Preventive Maintenance Manual
8 Copyright © 2012 EVG
3.2 Main Pressure
The Main Pressure is permanently checked by the software. There
must be an error message if the pressure drops below 5 bar.
Inspect the main pressure regulator (1) on the rear wall. The normal
static operating pressure should be around 6-8 bar.
Figure 1 - Main Pressure Regulator
Weekly check the main pressure regulator and the filter for the
presence of oil or water (see chapter “Weekly Maintenance”).
1
215 of 370