Weekly Maintenance
Copyright © 2012 EVG 63
4
4
5) Open Cover and insert Calibration Tool in opening. (Second
person shout hold the Calibration Tool/sheet with soft tension.
Otherwise Calibration tool/sheet could be crumpled)
6) Close bond cover.
7) Adjust calibration sheet to marking 1.
4.6.1.1 Check correct adjustment
Set the stack height to 500µm.
Activate wafer bow and pull out the paper strip. It should be possible
to pull it out slightly.
When setting stack height to 450µm, the paper should be clamped,
at 550µm it should be loosen.
270 of 370